GE Healthcare · CTLightSpeed230009041200352903
LightSpeed 4.X - Octane
Symptom
Power up recon data path self test fails System reconstructed scan data just fine Power up recon data path self test fails System reconstructed scan data just fine
Error message
Error: 230009041 The image did not complete reconstruction within the time allotted. See long description for complete error Tue Apr 15 19:45:47 2003 Error: 230009041 Host: smopct1 Process: queuemgr File: qmgr_queue_cntl.c Line: 2524 Function: Image Generation : SBC Processing Job Type: Retro Axial Scan: 414/4/1 Image: 414/4/5 The image did not complete reconstruction within the time allotted. Error: 200352903 Reconstruction errors detected See Long Description for error. Fri Jul 29 20:27:00 2005 Error: 200352903 Host: whct Process: reconSelftest File: reconSelftest.c Line: 200 Reconstruction errors detected Had Recon Retry on image 414/4/1, the recon job did not succeed errors in 1 images during selftest --------------------------------------------------------------------------- Fri Nov 17 12:32:05 2006 Error: 200352903 Host: ct04 Process: reconSelftest File: reconSelftest.c Line: 201 Reconstruction errors detected Had [IG 1] image checksum mismatch errors in 4 images during selftest
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2197229-5 ETHERNET SWITCH, 8 PORT used in 1X/2X/3X/4X GOC1/GOC2 See long description for more details Part #: 2197229-5 Part Type: GP CONSUMABLES Part Weight: 1.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ETHERNET SWITCH, 8 PORT FRU: Yes Status: Orderable Comments: FLS - ETHERNET SWITCH, 8 PORT. ADDEDPER PCN 220607. REPLACES P/N 2197229-3 AND -4. USE EXISTINGSTOCK FIRST. MS20544, 8/15/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.