Pular para o conteúdo principal
GE Healthcare · CTHiSpeed1831825000144858

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 488 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 44858/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 18:11:57:355 Log Series: 807 I/O board pilot relays are closed. PDU pilot relay not energized. BCSTAT = 0 HVDC Rail voltage = 210V If the rail voltage is approximately 200 Volts, you are missing one phase and most likely, a blown fuse is the culprit. HVDC Rail voltage = somewhere near 100 V This in-between voltage probably indicates bad C1 C2 capacitors or 3 phase Bridge Rect. or other PDU Control Bd. probl

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2139289-2 CPDU Control Bd. Replaces part 2139289 Used In CPDU - 2269902. 2139289-2 CPDU Control Bd. for Compact PDU Part #: 2139289-2 Part Type: GP REPAIRABLES Part Weight: 5.35 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 2139289-2 : PDU CONTROL BOARD W/ REW FRU: Yes Status: Orderable Comments: FLS - PDU CONTROL BOARD REWORK ASSEMBLY. THIS P/N 2139289-2, IS P/N T3202KF (2139289) WITH A REWORKED SUB-ASSEMBLY. ADDED PER PCN 220525 BY MS20544 ON 7/17/02. 9/10/02- 2139289-2 REPLACES 2139289 PER PCN 220525. MS12464 11/08/04 SSO# 212001261 FE's are to return defectives and unused/consigned back via FED Ex over night Item #16 A6 panel is the control board
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues