Pular para o conteúdo principal
GE Healthcare · CTHiSpeed186524186442186432186117

"Scan hardware Stopped Scan" with collimator errors

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 186524 Unexpected length or checksum of command was detected See long description for more details Sun Jul 18 07:27:42 2004 Error: 186524 Host: PCT2 Process: tgp_out_genetest File: tgp_errors.c Line: 312 DAS Error. Code : 30-1010-07 Error detected by : CIF Unexpected length or checksum of command was detected during communication between CIF and OGP.

Error: 186442 OGP did not receive any reply from CIF in response to SysConfig. See long description for more details Sun Jul 18 07:27:40 2004 Error: 186442 Host: PCT2 Process: tgp_out_genetest File: tgp_errors.c Line: 395 OGP Event Information. Code : 20-0030-07 Information detected by : OGP OGP did not receive any reply from CIF in response to SysConfig.

Error: 186432 OGP did not receive READY status from CIF See long description for more details Mon Jul 19 12:14:47 2004 Error: 186432 Host: PCT2 Process: tgp_out File: tgp_errors.c Line: 325 Aperture Error. Code : 20-1020-04 Error detected by : OGP OGP did not receive READY status from CIF for z-axis collimator control or OGP detected Aperture is not ready

Error: 186117 Error was detected at TGP_COLLIMATOR_REQUEST_RESP.(code=-1) See long description for more details Mon Jul 1

Applies to

Models
HiSpeed
Product
HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
Component
2271561-4 P2014XD CIF ASSY if das type is 2239788 on the Rating plate See the Long Description for a component location jpg 2271561-4 P2014XD DAS CIF ASSY if das type is 2239788 on the Rating plate for NXi systems Part #: 2271561-4 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: P2014XD CIF ASSY FRU: Yes Status: Orderable FLS - P2014XD CIF ASSY ADDED VIA SPCN CT-2069NP-2. REPLACES 2271561-3. MS26561 DSM 10/16/01 USED ON NXI
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues