GE Healthcare · CTLightSpeed
LightSpeed 4.X - GOC2
Symptom
Applications software fails to startup. After Linux boot up, output on both monitor screens stop with a login prompt displayed in top left corner of screen. Applications software fails to startup. After Linux boot up completes, the output on both monitor screens stop with a login prompt displayed in top left corner of screen. If you login and attempt to startup applications software it will fail to startup and reply back with: /usr/X11R6/bin/xterm Xt error: Can''t open display: :0.0
Error message
Error: Can''t open display: :0.0 (message seen on screen, not in an error log) From screen capture while booting the system. {ctuser@mhjx-gelsctb}2 st Waiting for kernel to clean up sockets... Error: Can't open display: :0.0 Sockets cleaned up. /usr/X11R6/bin/xterm Xt error: Can't open display: :0.0 cat: startMon.oc.status: No such file or directory *********** starting the HiSpeed CT/i system *********** *********** CT LightSpeed processes starting *********** {ctuser@mhjx-gelsctb}3 /usr/X11R6/bin/xterm Xt error: Can't open display: :0.0
gdm[1770]: gdm_slave_xioerror_handler: Fatal X error - Restarting :0 (error seen in /var/log/messages)
gdm[1749]: Failed to start X server several times in a short time period; disabling display :0 (error seen in /var/log/messages)
gdm[1749]: deal_with_x_crashes: Running the XKeepsCrashing script (error seen in /var/log/messages)Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405L.5_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console (2378862) 405L.5_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console (2378862) 405L.5_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console (2378862) Supports GOC2/ GRE LightSpeed 16 (4.X) Obsoleted by FMI 25338
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.