Pular para o conteúdo principal
GE Healthcare · CTLightSpeed244587244584178570260001008

LightSpeed 4.X - GRE

Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 244587 following subsystems failed to report initialization status in 50 seconds OBC SS(Coll/Das). Error: 244587 Host: qxi16 Process: fwmgr File: HwSetUp.c Line: 2095 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The following subsystems failed to report initialization status in 50 seconds: OBC SS(Coll/Das).

Error: 244584 Collimator/DAS reset failed See Long Description for complete error Error: 244584 Host: qx-ct Process: fwmgr File: HwManager.c Line: 1536 Collimator/DAS reset failed.

Error: 178570 Coll/Das controller reports failure in initializing tasks See Long Description for complete error Mon Mar 28 03:39:48 2005 Error: 178570 Host: ct04 Process: fwmgr File: HwSetUp.c Line: 3178 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 OBC_SS timeout: Coll/Das controller reports failure in initializing tasks.

Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long Description to see error example Mon Mar 28 03:39:48 2005 Error: 260001008 Host: OBC Process: HOP File: hop_node_init.c Line: 120 Functio

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2106993-5 FUSE 20 AMPS 500 VOLTS .3 SECONDS See Long Description for details Part #: 2106993-5 Part Type: GP CONSUMABLES Part Weight: 0.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 20 AMPS 500 VOLTS .3 SECONDS FRU: Yes Status: Orderable Comments: FLS - 20AMPS 500V .3SECOND FUSE U/O COST REDUCED GENERATORS SCPU 65/80 MS12746 PER RSPL 5-97 GELS - MAJ LEAD-TIME LE 99/07/26 PA,FUSE 20A 500V .3 SECONDS
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.