GE Healthcare · CTLightSpeed2600024033966726000245550723260001023260002358
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260002403 The Data Acquisition System Has Detected An Error In The -5V or +5V Analog DAS Power Supply. See Long Description for the full error code Tue Dec 7 07:20:51 2004 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 783 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:20:50:240 Log Series: 23 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply. Measured: 508 mV, Expected range: -5250 to -4750 mV Fri Jul 17 16:00:53 2009 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 729 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 39667/9/1 Type: Helical Exception Level: Sec/Soft Time: 16:00:54:485 Log Series: 357 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply.^B Measured: 4732 mV , Expected range: 4750 to 5250 mV Error: 260002455 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ) See Long Description for the full error code Tue Dec 7 07:23:22 2004 Error: 260002455 Host: DAS Process: DCB Firmware File: scanrx.c Line: 290 Fun
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225214 +/- 5V Analog H2 SDAS power supply includes the mounting plate, input/ output harness and power supply - used on 2.x and QXi See Long Description for details Part #: 2225214 Part Type: GP CONSUMABLES Part Weight: 5.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: POWER SUPPLY ASM CC CASE FRU: Yes Status: Orderable Parts Offered as: Current: 2225214 Harvested: Not Available Exchange: Not Available Comments: CT LIGHTSPEED H2 GANTRY 5V 6A ANALOG SUPPLY If no 2225214 power supplies are available part number 2218158 (H1 Gantry S-DAS) can be used, but the mounting bracket must be reused off the old power supply.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.