GE Healthcare · CTHiSpeed182904182912
Gantry table controls would not work when moving patient into the gantry.
Symptom
Gantry table controls would not work when moving patient into the gantry. Footpedals would move table up/down.
Error message
Error: 182904 A table side key is either stuck or pressed during initialization or just after Emergency Off released. See the Long Description for the full error code Thu Sep 16 06:48:08 2004 Error: 182904 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 345 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:48:00:070 Log Series: 20 A table side key is either stuck or pressed during initialization or just after Emergency Off released. Port value 0x0410. Disabling those keys detected being stuck Error: 182912 Gantry/Tableside FAST button is non-functional See the LD for the full error code Error: 182912 Gantry/Tableside FAST button is non-functional Thu Sep 16 06:48:08 2004 Error: 182912 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 406 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:075 Log Series: 20 Gantry/Tableside FAST button is non-functional
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required