GE Healthcare · Nuclear MedicineInfinia

L-mode fails step and shoot past 180 degrees with error 167 Max Dac on rotate axis

Symptom

L-mode fails step and shoot past 180 degrees with error 167 Max Dac on rotate axis L-mode fails step and shoot past 180 degrees with error 167 Max Dac on rotate axis. Error on rotate axis for step and shoot only, HHC and continous tomo rotate motion do not fail L-mode fails step and shoot past 180 degrees with error 167 Max Dac on rotate axis

Error message

Error #144 (Error 144) Velocity test fail axis = $ Axis # Axis Name 0 Lateral 1 Radial #1 2 Radial #2 3 N/A 4 N/A 5 Swivel #1 6 Swivel #2 7 Rotation 8 Linear 9 Height

Error #23 (Error 23) Software failure send buffer

Error #167 (Error 167) Max Dac Command Axis = $ Axis # Axis Name 0 Lateral 1 Radial #1 2 Radial #2 3 N/A 4 N/A 5 Swivel #1 6 Swivel #2 7 Rotation 8 Linear 9 Height

Applies to

Models
Infinia
Product
VGH VG and VG Hawkey and Varicam
Component
B010200028 VG rotation motor B010200028 VG rotation motor Part is obsolete see ITF 5408614-1EN for ordering part UGP001489 - Infinia SAX driver. 5409135 - VG rotation motor SP, this is the Infinia motor but with longer cables, 2 m instead of 1 m. 5408499 - set of brackets and transformer required for replacement of Bru 200 by SAX. Upon releasing of the parts Haifa will release also a Service Note. *Gantry Gantry Gantry
Area
Nuclear Medicine

Full cause & step-by-step resolution