Asset Scan Wont Run Due To Different Scan Range Than Asset Cal Scan.
Symptom
Will Not scan ASSET The only sequence that fails in this system is when the customer scans with the ASSET option.
Error message
2218527 Scanning has stopped. Error, please try again 2218527 lx-mr scn MsaScanOpsScrn.m 916 Scanning has stopped. error please try again 2227990 No valid ASSET calibration was found. 2227990 File: APS:ed_rdb_hdr_check Line: 1 No valid ASSET calibration was found. An ASSET scan must have the same coil, position, entry, and landmark as the most recent ASSET calibration.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution