GE Healthcare · MRExcite2245100741345574386

ACGD Gradient Driver Test Fails Due to the SGA ACGD Power Supply.

Symptom

ACGD Gradient Driver Static Test Fails ACGD Gradient Driver Test Fails

Error message

2245100 GRADIENT DRIVER TESTS - Static Test Failed on % Axis Fri Jun 11 17:57:56 2004 Error: 2245100 Host: lx-mr Process: diagCoord File: grad_lib.c Line: 416 Diagnostic: GDT Stat&Dyn Iteration: 1 of 1 Action: 1 Task ID: 0xfc9990 Time: 741345 Object: SCP Test Task GRADIENT DRIVER TESTS - Static Test Failed on X A Fri Jun 2 17:37:01 2006 Error: 2245100 Host: lxmr3 Process: diagCoord File: grad_lib.c Line: 921 Diagnostic: GDT Static Iteration: 1 of 1 Action: 1 Task ID: 0x3d5c6c8 Time: 574386 Object: SCP Test Task GRADIENT DRIVER TESTS - Static Test Failed on the Z Axis. VBUS_700 had 2 failures out of 39. Ready Static Mode-> this happened on the z axis- tech swapped the power inputs and the error moved-ordred SGA power supply

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required