GE Healthcare · CTLightSpeed260001085245206
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Can't scan with Pink Warning: Software has detected an invalid DAS configuration.
Error message
Error: 260001085 Communication error: Message not sent to the DCB. Communications with the DCB are down. CAN message could not be sent. Message code: 11 See Long Description for complete error Sun Sep 5 08:05:00 2004 Error: 260001085 Host: OBC Process: HOP File: hop_config_track.c Line: 164 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:04:45:705 Log Series: 30 Communication error: Message not sent to the DCB. Communications with the DCB are down. CAN message could not be sent. Message code: 11 Error: 245206 Software has detected an invalid or mistached DAS type. This will prevent t he system from scanning. Sun Sep 5 08:11:02 2004 Error: 245206 Host: ct01 Process: scanmgr File: validate_request.cxx Line: 341 Software has detected an invalid or mistached DAS type. This will prevent t he system from scanning.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225214 +/- 5V Analog H2 SDAS power supply includes the mounting plate, input/ output harness and power supply - used on 2.x and QXi See Long Description for details Part #: 2225214 Part Type: GP CONSUMABLES Part Weight: 5.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: POWER SUPPLY ASM CC CASE FRU: Yes Status: Orderable Parts Offered as: Current: 2225214 Harvested: Not Available Exchange: Not Available Comments: CT LIGHTSPEED H2 GANTRY 5V 6A ANALOG SUPPLY If no 2225214 power supplies are available part number 2218158 (H1 Gantry S-DAS) can be used, but the mounting bracket must be reused off the old power supply.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.