GE Healthcare · CTLightSpeedVCT1824271203314660
LightSpeed 3.X (Ultra, Plus, QXi) - GRE
Symptom
Gantry tilts slowly 1 degree and stops - table moves very slowly 5mm and stops. Gantry tilts slowly 1 degree and stops - table moves very slowly 5mm and stops
Error message
Error: 182427 Gantry Back / Forward Touch Sensor Error See Long Description for complete Error Error: 182427 Gantry Tilt Back or Gantry Tilt Forward Sensor Error Error: 182427 Host: ETC Process: Elevation File: emcTouchSensor.c Line: 125 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 12033/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 14:58:23:525 Log Series: 344 The Gantry BACK Touch Sensor Error : Console Pushbutton Remote Tilt is DISABLED. Use Gantry Controls to Tilt the Gantry. Current Position 14660 (.01 mm). Elev Dir: 1 (0 = Up 1 = Down) (Elev Cmd: 1 TouchSensor: 0x8000). Same Error for the Forward Touch sensor
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required