57723 Scan HW Faults

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 245215 Scan setup failed on Step:7 (Get Cal Module). Status:-1 See Long Description for complete error Tue Oct 11 10:45:52 2005 Error: 245215 Host: er_ct5_vct Process: scanmgr File: scanmgr_lib.cxx Line: 1224 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 524000 Scan setup failed on Step:7 (Get Cal Module). Status:-1 Fri Apr 27 12:05:20 2007 Error: 245215 Host: jrct Process: scanmgr File: scanmgr_lib.cxx Line: 1154 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 2910000 Scan setup failed on Step:3 (Allocate Scan File). Status:245047

Error: 230008017 The Scan File Manager server had an abnormal shutdown the last time that it ran, possibly due to power failure. See long description for more details Mon Dec 18 19:03:52 2006 Error: 230008017 Host: 77p_ct_exrm1 Process: scanfilemgr File: sfm_main.c Line: 1899 Function: System Control : Scan Database Manager The Scan File Manager server had an abnormal shutdown the last time that it ran, possibly due to power failure. Scan file consistency checks will be performed.

Error: 230019006 A non-zero errno was encountered during a remote system call. See Long Descrip

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.