Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23000502423000653025088000000800000230013015

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Multiple recon retries and suspended images during an exam. See long description for details Multiple recon retries and suspended images during an exam. To get the images to recon you would have to resume the queue or reboot the system for the images to recon.

Error message

Error: 230005024 image recon seq id: 18997Rebin failed in row delay, status=2 See Long Description for complete error Mon Feb 6 12:12:50 2006 Error: 230005024 Host: ice-post Process: sdcctrl File: sdcctrl_events.cxx Line: 447 image recon seq id: 18997Rebin failed in row delay, status=2

Error: 230006530 RECON_JOB_FAILED_RETRY See Long Description for complete error Tue Feb 1 06:33:34 2005 Error: 230006530 Host: ice Process: reconmgr File: rm_recon_mgr.c Line: 1114 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 25088/2/9 Image: 25088/3/25 Recon Seq Id: 957 RECON_JOB_FAILED_RETRY : number_detector_macro_rows 4 collimation_mode 0 num _rows_used_per_image 1 vw_mode -1 zsmooth_factor 1.000000 ibo 1 tilt 0.000000 gantry_period_sec_rev 0.800000

Error: 230013015 xx DAS views of 100 retrieved from the scan database failed checksums See Long Description for great info on this error Problem: The Following errors all point to a scsi Scan data disk issue. Using the gesys_Ct01.log and iceConsole.log is a must, together both logs point to the Scan data disk chain. Check DIP stats if stats are clean the das data is good. The iceConsole log is a little difficult to r

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 Class C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395177 2395162, 2395163, 2399867 LightSpeed Plus/ Ultra MDAS Apps Software (Disk 1, Disk 2 and Patch A) 308I.2_H3.1M5 FW47-03 LS2002M5. IRIX Only. Class C/M Common w/ LS 2X, 3X, & 4X.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues