GE Healthcare · MRExciteSigna22488222248850
Auto Prescan failed, RF Amplifier is not ready
Symptom
Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.
Error message
2248822 Analogic RF Amplifier 1 High Voltage Low Fault Mon Nov 27 10:03:56 2006 Error: 2248822 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 Analogic RF Amplifier 1 High Voltage Low Fault. Analogic Amplifier 1 power supply voltage dropped below 120V DC.(150V nominal).-> 2248850 RF Amplifier fault 71 Mon Nov 27 10:10:04 2006 Error: 2248850 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 RF Amplifier fault 71.
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.