Array processor (AP) errors during scanning Due to Master AP Board .

Symptom

Array processor AP errors during scanning. Scans stop (abort) with an AP error logged. It appears that AP crashes are not related to heavy duty recon jobs or high data acquisition rates. A typical set of stress scans did not reproduce AP crashes. AP error file doesn't reveal any pattern either. Site currently has old style (Reflex-400A) boards. These boards were swapped for the newer Reflex-400B, but there was no change in symptoms. SPR MRIhc06568

Error message

2228010 AP Error

2228016 Internal AP software error See Long Description Please consult file /usr/g/service/log/aperror.log for more information.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2325947-12 AP Reflex Minerva II 400B Master Board - M3000RJ For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering.
Area
MR