GE Healthcare · MRSigna2225668
Unable to scan, Erbtec errors.
Symptom
Unable to scan with Erbtec errors
Error message
2225668 ERBTEC RF Amplifier fault 89. Tube air problems See Long Description Error: 2225668 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3060 ERBTEC RF Amplifier fault 89. Tube air problems. Possible defective blower, Power Control Board, sensor, and cables.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.