Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT230017174230017078503421111522307108580230017083

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Fastcal fails the mylar window check. Fastcal fails the mylar window check

Error message

Error: 230017174 ACCS: Air Calibration Refresh Calculation (BEAM OBSTRUCTION QUAD) failed. See Long Description for complete error Mon Oct 31 07:08:03 2005 Error: 230017174 Host: cpcvct2p Process: accs File: aircal_refresh_processing.c Line: 333 ACCS: Air Calibration Refresh Calculation(BEAM OBSTRUCTION QUAD) failed.

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information

Error: 230017083 Possible beam obstructed - sin/cos vector magnitude = 0.xxxxxx - may include row and channel numbers See LD for complete error Mon Oct 31 07:08:03 2005 Error: 230017083 Host: cpcvct2p Process: accs File: aircal_spectral_acal_util.c Line: 445 Possible beam obstructed - sin/cos vector magnitude = 0.385870 Mon Mar 21 07:27:12 2011 Error: 230017083 Host: sj01_ct64s_6001 Process: accs File: aircal_spectral_acal_util.c Line: 449 Possible beam obstructed - row 58(26B), detector channel 375, sin/cos vector magnitude =

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues