GE Healthcare · CTLightSpeed2600010122600040005004226000104911497260001071
LightSpeed 4.X - GRE
Symptom
Scanning aborted midway through exam with scan communication errors.
Error message
Error: 260001012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted See Long Description for the full error code Sat Jun 9 14:37:06 2001 Error: 260001012 Host: OBC Process: HOP File: hop_fault_state.c Line: 137 Function: No System Function Reported Scan: 9553/3/1 Type: Axial Exception Level: Pri/Most Severe Time: 14:37:06:030 Log Series: 366 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. None of the sub-systems have reported a problem. Check nodes or fault line for failure. Error: 260004000 DAS firmware object: Lost communications with controller See Long Description for more details Wed Nov 22 05:44:35 2006 Error: 260004000 Host: OBC Process: DAS File: ss_comm.c Line: 329 Function: SYSTEM COMMUNICATIONS : Rotating Controller I/F Bus Scan: 50042/6/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 05:44:35:175 Log Series: 81 DAS firmware object: Lost communications with controller Error: 260001049 Communications with the DCB sub-system has not been established. Scanning aborted. See LD for complete error Wed Nov 22 07:15:07 2006 Error: 260001049 Host: OBC Process: HOP File: hop_misc.c Line: 265 Function: No System Functi
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2318704 DCB MDAS 16 slice See LD for details Part #: 2318704 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -14549.85 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 16 SLICE DCB MDAS FRU: Yes Status: Orderable Parts Offered as: Current: 2318704 Harvested: Not Available Exchange: Not Available Comments: FLS - 16 SLICE DCB MDAS ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.