GE Healthcare · CTLightSpeedVCT
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
HP Host computer locks up HP 8000 or HP 8200 computer locks up
Error message
HP 8000 Error: Failed to start X server several times in a short time period; disabling display :0 See LD for more details [root@emc1]# cd /var/log [root@emc1]# cat messages | grep server Oct 30 17:12:03 emc1 gdm[3422]: Failed to start X server several times in a short time period; disabling display :0 Oct 30 17:31:28 emc1 gdm[3460]: Failed to start X server several times in a short time period; disabling display :0 Oct 30 17:32:36 emc1 kernel: nfsd: last server has exited Oct 30 17:52:00 emc1 gdm[3539]: Failed to start X server several times in a short time period; disabling display :0 Oct 30 18:45:43 emc1 gdm[3473]: Failed to start X server several times in a short time period; disabling display :0 Oct 30 18:55:07 emc1 gdm[3478]: Failed to start X server several times in a short time period; disabling display :0
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2368697-2 LINUX 3 BUTTON PS2 MOUSE See LD for details Part #: 2368697-2 Part Type: GP CONSUMABLES Part Weight: 0.22 LBR Repairable: No Returnable: No Hazardous Material: N Description: LINUX 3 BUTTON PS2 MOUSE FRU: Yes Status: Orderable Comments: FLS - LINUX DDR WORKSTATION 3 BUTTON PS2 MOUSE . THIS P/N REPLACES P/N 2330676-2 FOR THE GOC2-LINUX CONSOLE, PER PCN 220795. MS20544 ON 2/19/03. DO NOT!!!! CHANGE TO 2368697 THIS IS THE WHOLE COMPUTER! IN CASE YOU MAY NEED IT, YOU CAN PROCURE INSTEAD 2348405-2. (HP P/N Y1755A). THIS HAS BEENIDENTIFIED TO BE THE SAME AS P/N 2368697-2 (HP P/N A28314).(THE DIFFERENT HP P/NS ARE DUETO INTERNAL SUPPLY CHAIN INTEGRATION BETWEEN HP AND COMPAQ).MS20544 ON 3/11/03. PART PLACED ON CONSUMABLE COLLECTIO PER REQUEST OF CHAD VANDE HEI MS13452 030519 ALSO USED ON EXCITE II FROM MR, PER RSPL. MS20544 6/17/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.