GE Healthcare · CTLightSpeed260100028
LightSpeed 5.X (60kW)
Symptom
Unable to do axial or helical scans
Error message
Error: 260100028 Scan Control timed out waiting for the Ready For Views message from the SRU. See LD Fri Jan 7 09:25:36 2005 Error: 260100028 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 1539 Fri Jan 7 09:25:35 2005 Host : Table Gantry Processor Ermes # : 260100028 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 1539 Function : DATA ACQUISITION : Scan Data Acquisition Scan Type : Scout Scan: 470/1/2 Scan Control timed out waiting for the Ready For Views message from the SRU. This message is sent from the SRU once offset collection has comple ted. Possible causes: trigger failure, DIP failed to receive data, message lost, software error.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2379129 scsi cable between the SDDA and the DARC See Long Description for Details. Part #: 2379129 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: GRE CONSOLE, SCSI CABLE KIT FRU: Yes Status: Orderable Comments: FLS - GRE CONSOLE, SCSI CABLE KIT USED ON GRE / H16 ADDED PER GRE RSPL MS13452 030905 RISK: FUNCTIONALITY FREQUENCY: LOW CONTAINS CABLE TO BE REWORKED MS13452 FOR SCREWLOCK ISSUE. This is NOT just a Scsi cable, but it is a Scsi cable kit containing the following cables GOC3, SCSI Cable Kit 2379129 Part #: 2379129 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: GRE CONSOLE, SCSI CABLE KIT FRU: Yes Status: Orderable The kit contains the following cables 1 2324697-2 1.000 Yes SCSI CABLE, VHD68 TO CENT 50, LINUX 2 2323942 1.000 Yes CABLE, HD50 TO HD50 3 2376295 1.000 Yes CABLE, SCSI U320 HD68M/M 305MM
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required