Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT1831803300318328950001183108

LightSpeed 1.X with S-DAS4 Gantry and GRE

Symptom

Scanning aborted with unexpected Loss Of High Voltage.

Error message

Error: 183180 I/O board fault detected. Relay K245 open, HVDC Bus not active. Cannot close backup contactor. EEBCSTAT = 1 KILLBC = 0 See Long Description for complete error Sun Jun 5 23:17:07 2005 Error: 183180 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 450 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:06:735 Log Series: 425 I/O board fault detected. Relay K245 open, HVDC Bus not active. Cannot close backup contactor. EEBCSTAT = 1 KILLBC = 0.

Error: 183289 Unexpected Loss Of High Voltage Wed Aug 4 20:21:15 2004 Error: 183289 Host: OBC Process: Generator File: gen xray.c 1.31 Line: 528 Function: XRAY GENERATION : Rotating HVON Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:21:16:325 Log Series: 59 Unexpected Loss Of High Voltage. .

Error: 183108 KV fault detected: Cathode shoot-thru. See long description for full error Sun Jun 5 23:17:05 2005 Error: 183108 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 248 Function: XRAY GENERATION : KV Loop Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:05:545 Log Series: 424 KV fault detected

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P92 SEMICONDUCTOR FUSE - 60A 700V Part #: 46-170021P92 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SEMICONDUCTOR FUSE FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60 FLS - FE ASSIST: IN ORDER TO INCREAS INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCRESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE U PART WHEN AVAILABLE. IF NO U USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. SEMICONDUCTOR FUSE FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues