Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011822226011821550368

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118222 (60-0401) No mA feedback. mA measurement function: PPC board has detected no mA feedback 20 ms after the beginning of the exposure.

Error: 260118215 (60-0318) kV did not reach 75 percent after 20ms See Long Description for full error Sat Mar 25 15:35:28 2006 Error: 260118215 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sat Mar 25 15:35:27 2006 Host : Jedi Ermes # : 260118215 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50368/1/3 (60-0318)kV did not reach 75 percent after 20ms. Indicates that there was no HV ON signal generated for exposure time count-up.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3634T (aka 2378211-2 ) Varian PERFORMIX PRO TUBE used on : HP80 AND LS RT4 See Long Description for details Part #: D3634T Part Type: GP TUBES Part Weight: 176.37 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX PRO (VARIAN) TUBE UNIT FRU: Yes Status: Orderable Comments: FLS 11/19/04 (Shawn Fox 212001261) PLEASE remember to complete the Tube report for the tube you are replacing and return it with old tube. The blank Tube report comes with the new tube. If you have any questions, please contact Yolanda Scheper 262-548-2864. If you do not return tube report with old tube, warranty credit can be missed or delayed for your Zone
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.