Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18260751200

Gantry tilt moves a small distance then stops (less than 2 deg) Gantry tilts forward a small distance then sto…

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Gantry tilt moves a small distance then stops (less than 2 deg) Gantry tilts forward a small distance then stops

Error message

Error: 182607 HW/SW error: Tilt Velocity tracking See LD for more details Thu May 11 23:04:54 2006 Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/95 Line: 139 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 51200/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 23:04:53:540 Log Series: 144 HW/SW error: Velocity tracking error. Actual Velocity 150 (.01 deg/sec) Expected Velocity 100 (.01 deg/sec) Tol 40 (.01 deg/sec) Cur Pos -1194 (.01 deg) Tilt Dir: 1 (0 = BCK 1 = FWD) . INFO: Probable Problem could be 180V Power is LOW.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues