GE Healthcare · CTLightSpeed200002352200501502
LightSpeed 5.X (80/100kW)
Symptom
Software shut itself down after clicking confirm to start scan series Software shut itself down after clicking confirm to start helical scan series
Error message
Error: 200002352 A HOST_SHUTDOWN was initiatied by the death of examRxEnv. See LD for complete error Wed Apr 5 12:02:45 2006 Error: 200002352 Host: gect_16 Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiatied by the death of examRxEnv. Error: 200501502 UifXIoErrorHandler: Encountered an X library I/O error. errno = xx Process will intentionally dump core. See Long Description for a complete error. Tue Nov 8 12:03:08 2005 Error: 200501502 Host: ct03 Process: environmentMgr File: UifXErrorHandlers.c Line: 124 Function: No system function reported UifXIoErrorHandler: Encountered an X library I/O error. errno = 32 Process will intentionally dump core.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area