Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230017801230017812000500260007419

LightSpeed 5.X (80/100kW)

Symptom

Collimator Cal fails in fastcals. Collimator Cal fails in fastcals

Error message

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Error: 230017812 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 See Long Desciption for complete error Mon May 30 05:41:49 2005 Error: 230017812 Host: cxi1 Process: accs File: collSweepProcessing.c Line: 491 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500

Error: 260007419 Data Acquisition System Temperature Is Low. See the LD for the complete error code Error: 260007419 Data Acquisition System Temperature Is Low. Mon Feb 6 08:37:56 2006 Error: 260007419 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 474 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 1277 Type: Axial Exception Level: Pri/Soft Time: 18:27:37:070 Log Series: 20 Data Acquisition System Temperature Is Low

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2404393 GDAS-16 CONVERTOR CARD, replaces 2364354. See Long Description for details. 2404393 GDAS-16 CONVERTER CARD, replaces 2364354 Part #: 2404393 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: 16 SLICE GDAS CONVERTER CARD FRU, A CONVERTER CARD IN AN ESD BAG IN A BOX WITH A PART NUMBER LABEL FRU: Yes Status: Orderable Comments: FLS - 16 SLICE GDAS CONVERTER CARD USED ON: LIGHTSPEED PRO 16 FRU GDAS ADDED PER GDAS RSPL MS13452 040315 RISK: SITE DOWN FREQUENCY: HIGH NOTE: THERE ARE 96 OF THESE CARDS PER SYSTEM, BUT SHOULD NOT NEED TO REPLACE ALL AT ONETIME, MS13452 THIS PART CONTAINS ONE GDAS CONVERTER CARD 2364354 WITH FRU PACKAGING, SUPPLIED BY GEMS IT. MS13452 PDR 8471.80.9000 IM05 UPDATE AWAITING MM (212001261 4/1/06) For 4/8 slice G Das which uses a thick connector must order 2364354-2
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.