GE Healthcare · CT18315130948183158
A loud pop or bang was heard from the gantry.
Symptom
A loud pop or bang was heard from the gantry. The customer was unable to, complete the scan or scan any more patients. A loud pop or bang was heard from the gantry. The customer was unable to, complete the scan or scan any more patients
Error message
Error: 183151 Broken stator wires detected. See LD for complete error Sun Oct 29 14:26:22 2006 Error: 183151 Suite: PNSL Host: OBC Proc: Generator File: ctvrc_manager.c 1.8 Method: No Method Line: 136 Function: XRAY GENERATION : Tube Rotor Control Scan: 30948/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 14:26:22:940 Log Series: 322 Broken stator wires detected. Green stator = 0.037 A Black stator = 0.055 A White stator = 0.043 A Error: 183158 HVDC Bus voltage not sensed in 2 seconds See LD for complete error Wed Sep 20 11:01:10 2006 Error: 183158 Suite: SMCT Host: OBC Process: Generator File: gen_rotor.c 1.40 Method: No Method Function: SYSTEM MONITORING : 550V Monitor Scan: 0/0/1 Type: Static Line: 530 Exception Level: Pri/Most Severe Time: 11:01:10:450 Log Series: 25 HVDC Bus voltage not sensed in 2 seconds. Pilot relays on I/O board closed, pilot relay in PDU energized.
Applies to
- Product
- HSA RP 1.X HSA RP 1.X with Zeus HSA Gantry and Sun Genesis Console FMI 25246 HSA/RP 1.6 Y2K Upgrade Special Tools Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P79 30A 700v Fuse in CTVRC See LD for Details Part #: 46-170021P79 Part Type: GP CONSUMABLES Part Weight: 0.02 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30 AMP/700V.FUSE FRU: Yes Status: Orderable Comments: FLS - THIS IS THE 30A 700V FUSE USED ON THE FILAMENT POWER ASSEMBLY OF THE CT HISPEED ADVANTAGE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.