GE Healthcare · MRExciteSigna2220008221863022185012224854

Coil ID Fails Due to 8 channel switch

Symptom

Coil ID Fails and CoilID Light is Red See Long Description Customer is attempting to scan, but getting error that something is wrong with coil ID. The Coil ID LED turns red.

Error message

2220008 RF coil connection failure detected, reconnect coil and if problem persists contact GE Service. 2220008 5 NSR334 scn RxMgrDownload.m 512 RF coil connection failure detected, reconnect coil and if problem persists contact GE Service. Upper coil connect status:1, Upper coil code:Null, Lower coil connect status:0, Lower coil code:Null

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

2224854 SRI 8 Channel Switch Comm Error. SRI Received Packet from the 8 Channel Switch which did not Include all Parameters. Wed Apr 14 08:32:03 2004 Error: 2224854 Host: exmr Process: NSP File: SCP:SRI Line: 350 SRI-8 Channel Switch Comm Error. SRI received packet from the 8 Channel Switch which did not include all parameters. If "Actual Cnt" is 0, fiberoptic cables between SRI and 8 Channel Switch may be disconnected. Cmd: 4 Expect Cnt: 5 Actual Cnt: 0

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2362995-2 Eight 8 Channel Switch
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.