Infinia Scan stoppage occured indicating to the customer that there has been a Hardware communication error.

Symptom

Infinia Scan stoppage occured indicating to the customer that there has been a Hardware communication error. Scan stoppage will occur indicating to the customer that there has been a Hardware communication error. The error log will have the following error” FE_WRITE_PORT_ERROR:.:Hardware communication error:.:Try again Perform system reset.:.: Failed to write Front End port: ( Port: 0x0 on POP board):.:Trace:{1}” Infinia Scan stoppage occured indicating to the customer that there has been a Hardware communication error. The error log will have the following error” FE_WRITE_PORT_ERROR:.:Hardware communication error:.:Try again Perform system reset.:.: Failed to write Front End port: ( Port: 0x0 on POP board):.:Trace:{1}”

Error message

FE_WRITE_PORT_ERROR:.:Hardware communication error:.:Try again Perform system reset.:.: Failed to write Front End port: ( Port: 0x0 on POP board):.:Trace:{1

FE WRITE PORT ERROR:.:Hardware communication error:.:Try again Perform system reset.:.: Failed to write Front End port: ( Port: 0xa1e on DOC board of detector #1):.:Trace:{1} FE WRITE PORT ERROR:.:Hardware communication error:.:Try again Perform system reset.:.: Failed to write Front End port: ( Port: 0xa1e on DOC board of detector #1 ):.:Trace:{1}

Applies to

Models
Infinia
Product
Infinia II Infinia II Acquisition System
Software
Infinia All Infinia All. Including Infina 1 and Infinia II. Infinia All. Including Infina 1 and Infinia II.
Component
5252414 FOX POP BOARD (For New CDL) 5252414 FOX POP BOARD (For New CDL) PCA000526 > PCA000547> 5252414 *Gantry Gantry Gantry
Area
Nuclear Medicine