Pular para o conteúdo principal
GE Healthcare · CTHiSpeed18394965001

HiSpeed CT/i 4.X with SGI Indigo Console, HiSpeed Advantage Gantry

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183949 Could NOT CLOSE Gantry (Axial) Loop Contactor See LD for full error CT01 Host: STC Proc: Axial Error: 183949 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor Another example: Tue May 24 08:41:21 2005 Suite: CT02 Host: STC Proc Axial Error: 183949 File: ax_error.c Method No Method 1114 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 65001/11/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:21:940 Log Series: 60 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e28. EN 401 SR 442

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-264806G1 STC AXIAL CONTROL BOARD See LD for details Part #: 46-264806G1 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: AXIAL CONTROL BOARD FRU: Yes Status: Orderable Comments: FLS - FE: READ SN T-1345. IF U455 ONAX BOARD IS BEING SHORTED, THECABLE MAY BE PINCHED. MS8365 4/95 HQ INV ONLY: PCN'S 161734 AND 161831 IMPLEMENTED. STOCK WASPURGED. 8/93 MBELL FLS - FE ASSIST: MS6588 MAY 2002 46-264806G1 = AXIAL BD FOR CT1.0 SEC. AXIAL ROTATION. THIS BOARD HAS THE PALS FOR 1.0 SECOND ROTATION. THE T3202BD =1.0 SEC AXIAL BD. THE T3202KD = .8 SEC AXIAL BD. THE ONLY DIFFERENCE BETWEEN THE T3202KD AND T3202BD ARE TWO PALS. THE T3202BD CAN BE UPGRADED TO THE T3202KD BY CHANGING PALS U211 AND U101. DO YOU STILL WANT TO ORDER THIS PART?? 46-264806G1 STC AXIAL CONTROL BOARD
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.