GE Healthcare · CTLightSpeed200002368200002352
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Applications software fails to boot up completely and automatically shuts itself back down. Applications software fails to boot up completely and automatically shuts itself back down
Error message
Error: 200002368 A set of processes are deadlocked waiting for each other to start. This deadlock has caused the system to shutdown! See Long Description for full error. Mon Oct 9 07:55:18 2006 Error: 200002368 Host: ct01 Process: cupMonitor File: ssProcStart.c Line: 997 A set of processes are deadlocked waiting for each other to start. "afQueueMgr, examRxDisplayMgr, sfm_archive, scanmgr, scanfilemgr, headermgr, coordinator, rtsmgr, reconSelftest, queuemgr, procICE, accs, vavQueueMgr, stcIf, sdm, scanRx, retroRecon, reconMgmt, readcfgServer, ppsproxy, piRxMgmt, mtran, examRxEnv, dms, visScript, ces, awStart, WLServer, ObjMgr" are the likely culprits. This deadlock has caused the system to shutdown! Error: 200002352 A HOST_SHUTDOWN was initiatied by the death of agent. See LD for a full error example. Mon Oct 9 07:55:18 2006 Error: 200002352 Host: ct01 Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the death of agent. Failed to set network config for Visibroker found in /var/adm/SYSLOG: /var/adm/SYSLOG: Failed to set network config for Visibroker Error posted in /var/adm/SYSLOG: Oct 9 07:55:16 5B:ct01 ** Failed to set network config for Visibrok
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-54 Octane PCI Ethernet board See LD details, picture 2169940-54 Octane PCI Ethernet board Part #: 2169940-54 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: OCTANE PCI ETHERNET CARD FRU: Yes Status: Orderable Parts Offered as: Current: 2169940-54 Harvested: Not Available Exchange: Not Available Comments: OCTANE 1 & 2 PCI ETHERNET CARDTHIS IS THE SAME PART AS 2169940-8; IF NONE AVAILABLE THEN OFFER UU VERSION. COMPONENT COLLECTION AT RDC PER P KOLVENBACH. CONSUMABLE COLLECTION STARTED ON 2/27/03 PER MIKE DIGGINS 7/11/03 - SWITCHED TO DEX PER GLEN SABIN. GELS - CCP, ETHERNET CARD
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.