GE Healthcare · CTLightSpeed183182500014485821408024294
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 488 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 44858/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 18:11:57:355 Log Series: 807 I/O board pilot relays are closed. PDU pilot relay not energized. BCSTAT = 0 HVDC Rail voltage = 210V If the rail voltage is approximately 200 Volts, you are missing one phase and most likely, a blown fuse is the culprit. HVDC Rail voltage = somewhere near 100 V This in-between voltage probably indicates bad C1 C2 capacitors or 3 phase Bridge Rect. or other PDU Control Bd. probl
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2139289-2 CPDU Control Bd. Replaces part 2139289 Used In CPDU - 2269902. 2139289-2 CPDU Control Bd. for Compact PDU Part #: 2139289-2 Part Type: GP REPAIRABLES Part Weight: 5.35 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 2139289-2 : PDU CONTROL BOARD W/ REW FRU: Yes Status: Orderable Comments: FLS - PDU CONTROL BOARD REWORK ASSEMBLY. THIS P/N 2139289-2, IS P/N T3202KF (2139289) WITH A REWORKED SUB-ASSEMBLY. ADDED PER PCN 220525 BY MS20544 ON 7/17/02. 9/10/02- 2139289-2 REPLACES 2139289 PER PCN 220525. MS12464 11/08/04 SSO# 212001261 FE's are to return defectives and unused/consigned back via FED Ex over night Item #16 A6 panel is the control board
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.