GE Healthcare · CTLightSpeed1824271203314660
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Table will not lower all the way down and gantry will not tilt
Error message
Error: 182427 Gantry Back / Forward Touch Sensor Error See Long Description for complete Error Error: 182427 Gantry Tilt Back or Gantry Tilt Forward Sensor Error Error: 182427 Host: ETC Process: Elevation File: emcTouchSensor.c Line: 125 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 12033/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 14:58:23:525 Log Series: 344 The Gantry BACK Touch Sensor Error : Console Pushbutton Remote Tilt is DISABLED. Use Gantry Controls to Tilt the Gantry. Current Position 14660 (.01 mm). Elev Dir: 1 (0 = Up 1 = Down) (Elev Cmd: 1 TouchSensor: 0x8000). Same Error for the Forward Touch sensor
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.