GE Healthcare · MRExcite2220007
Coil not plugged in properly.
Symptom
Coil ID Fails and CoilID Light is Red See Long Description Customer is attempting to scan, but getting error that something is wrong with coil ID. The Coil ID LED turns red.
Error message
2220007 The prescribed RF coil and detected RF coil do not match. See long description on reading/interpreting this type of error message Note: An example of this error message, with explanation, is provided below. Wed Mar 3 08:18:58 2004 Error: 2220007 Host: lx-mr3 Process: scn File: RxMgrDownload.m Line: 2074 The prescribed RF coil and detected RF coil do not match. 040303104912225 Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:1, Lower coil code:q8dCUW0PD2EaETIplAp90PLrACM6CW38, Rx coil name:MRI 8HRKNRR PA, Rx coil code:q8dCUWOPD2EaETIpIAp9oPLrACM6CW38 NOTE: Upper coil connector status refers to connector plugged into upper port at "dog house" (i.e. head coil or Med' Adv' QUAD KNEE ("chimney") coil). Lower coil connect status refers to all PA coils / any coil using the lower port at the "dog house". Coil code should match that of what was detected (in this case the zeros were config'd as O and o (oh's - upper and lower case), and the number one following "ETIp" was an "I").Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.