GE Healthcare · CTLightSpeed260003120500422600031212600031185017250066
LightSpeed 3.X (Ultra, Plus, QXi) - Octane
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003120 An error occurred attempting to home CAM A or B See LD for complete error Mon Jan 10 09:04:41 2005 Error: 260003120 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 193 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50042/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 09:04:41:350 Log Series: 57 An error occurred attempting to home CAM A. An error occurred attempting to home CAM B. Error: 260003121 The Aperture failed to move to the requested position. See LD for complete error Mon Dec 6 12:55:55 2004 Error: 260003121 Host: HSC Process: CCB Firmware File: aperture_move.c Line: 225 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:55:56:165 Log Series: 63 The Aperture failed to move to the requested position. Error: 260003118 CAM A drive tracking detected the current motor position is out of the tolerance band for a successfull move See Long Description for complete error. Error: 260003118 CAM A drive tracking detected the current motor position is out of the tolerance
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5404254 (replaces 2125453-3) Collimator cam motor driver amp - used on 1.X (QXI) 2.X, 3.X, 4.X and 5.X. Not used on VCT See LD for details 5404254 2125453-3 Collimator cam motor driver amp -used on 1.X (QXI) 2.X, 3.X, and 4.X. Not used on VCT Part #: 5404254 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Helios Collimator CAM Drive FRU: Yes Status: Orderable Part #: 2125453-3 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: HELIOS COLLIMATOR, CAM MOTOR DRIVER FRU: Yes Status: Orderable Comments: FLS - CAM MOTOR DRIVER HELIOS COLLIMATOR MS08992 9/3/98 M3 4TH QTR '98 HQT: VENDOR ORIENTAL MOTOR LEAD TIME WITH AIRFEIGHT 12 DAYS BY BOAT 6 WEEKS. INCREMENTAL COST FOR AIR FREIGHT $2
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required