Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182417182408182404

LightSpeed 4.X - Octane

Symptom

After a hardware reset, the table is raised and lowered through full range but will not reference. Flashings 8s would not go away in the elevation direction.

Error message

Error: 182417 Invalid Index Pulse with reference to the Interference switches Sun Sep 18 10:02:45 2005 Error: 182417 Host: ETC Process: Elevation File: emcMonitor.c 1.44 1/29/96 Line: 111 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:02:41:090 Log Series: 27 HW error: Invalid Index Pulse with reference to the Interference switches.

Error: 182408 Velocity tracking error. Probable Problem could be 180V Power is LOW. See Long Description for complete error Error: 182408 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 83 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 11:57:28:685 Log Series: 35 HW/SW error: Velocity tracking error. Actual Velocity 0 (.1 mm/sec) Expected Velocity 400 (.1 mm/sec) Tol 320 (.1 mm/sec) Cur Pos 3276 (.1 mm) Elev Direction: 1 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.

Error: 182404 HW error: No Index Pulse between Interference switches. See LD for error as seen in viewlog Error: 182404 HW error: No Index Pulse between Interference switches Wed Aug 13 08:05:33 2003 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269173 ELEVATION ENCODER ASM FOR H2 LCC -Zeus Table see Long Description for details and picture Part #: 2269173 Part Type: GP CONSUMABLES Part Weight: 3.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER ASM FOR H2 LCC FRU: Yes Status: Orderable Comments: FLS - ELEVATION ENCODER ADDED. USED ON H2 TABLE LCC. ASIA HAS IDENTIFIED THIS AS A LINKED PART (SEE 46-296633G3).AWAITING CONFIRMATION ON GL18.MSA8429 11/8/01 WILL START DEPLOYMENT FROM ASIA ON FW12, 02. MS30544 1/31/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues