Cannot scan in any transmit mode due to Head and Body Path Open and Shorted T/R Driver Faults.

Symptom

Cannot scan in any transmit mode due to Head and Body Path Open and Shorted T/R Driver Faults. All of the error codes 2225517, 18, 19, & 20 are logged on every scan attempt no matter if it is tried in body or head transmit mode. FE found "smoked" components on the CPD board inside the SSM. But after replacing that board and recalibrating it, the problem persists. FE gets diode action when measuring the T/R lines.

Error message

2225520 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load. See Long Description Thu Sep 9 11:16:03 2004 Error: 2225520 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6080 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.