Cannot scan in any transmit mode due to Head and Body Path Open and Shorted T/R Driver Faults.
Symptom
Cannot scan in any transmit mode due to Head and Body Path Open and Shorted T/R Driver Faults. All of the error codes 2225517, 18, 19, & 20 are logged on every scan attempt no matter if it is tried in body or head transmit mode. FE found "smoked" components on the CPD board inside the SSM. But after replacing that board and recalibrating it, the problem persists. FE gets diode action when measuring the T/R lines.
Error message
2225520 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load. See Long Description Thu Sep 9 11:16:03 2004 Error: 2225520 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6080 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.