Pular para o conteúdo principal
GE Healthcare · CTHiSpeed1831825000144858

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 488 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 44858/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 18:11:57:355 Log Series: 807 I/O board pilot relays are closed. PDU pilot relay not energized. BCSTAT = 0 HVDC Rail voltage = 210V If the rail voltage is approximately 200 Volts, you are missing one phase and most likely, a blown fuse is the culprit. HVDC Rail voltage = somewhere near 100 V This in-between voltage probably indicates bad C1 C2 capacitors or 3 phase Bridge Rect. or other PDU Control Bd. probl

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296221P15 80A 600V Fuse See Long Description for details Part #: 46-296221P15 Part Type: GP CONSUMABLES Part Weight: 0.13 LBR Repairable: No Returnable: No Hazardous Material: N Description: BUSSMANN KBC80 FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU. PDV BUSSMANN KBC80 80A, 600V THERE ARE 3 OF THESE FUSES IN FUSE KIT 2218570 - MS4751 8/01 PRICE CHANGE PER BRIAN NAGLE 4/98 FLS - FE ASSIST: IN ORDER TO INCREAS INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCRESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE U PART WHEN AVAILABLE. IF NO U USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues