GE Healthcare · CTHiSpeed219800183327183334

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scanner Hardware Stopped Scan HEMRC Errors See Long Description for more information. Scanner Hardware Stopped Scan message appears during Helical Scanning with the following Errors: 185553 (The HEMRC drive has detected a Line Loss - Alarm status: 20H. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage) Possible causes: X-Ray tube stator , HEM-IT, HEMRC power supply 183318 (The HEMRC output current is out of 45.0 percent WARNING tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.097 A. 183319 (The HEMRC output current is out of 55.0 percent MAXIMUM tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.653 A. 183316 (The HEMRC failed to reach the Run speed in the allotted time), and 185548 (The HEMRC operating frequency is below the minimum value) Actual frequency: 106.973 Minimum frequency limit: 130.000 At Frequency signal: Inactive Expected: Inactive Rotor State = Transition

Error message

Error: 219800 The HEMRC AC Drive reported fault code: F2 - Auxiliary Fault See LD for complete error Error: 219800 The HEMRC AC Drive reported fault code: F2 - Auxiliary Fault Wed Sep 20 10:47:58 2006 Error: 219800 Suite: SMCT Host: OBC Process: Generator File: HEMRClogFault.c 1.2 Method: No Method Line: 196 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:47:53:665 Log Series: 23 The HEMRC AC Drive reported fault code: F2 - Auxiliary Fault The interlock between the Chopper Control circuit on the HEMRC I/F Board and the Drive is open. Possible Chopper Control fault or connections to the I/F board. Also check the fuse on the chopper resistor pan.

Error: 183327 HEMRC fault detected. Long Description contains complete error. Mon Jul 24 08:41:32 2006 Error: 183327 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 625 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:28:555 Log Series: 31 HEMRC fault detected. See system log for additional error information. Fault latch address = ffb82b hex, Bit = D0

Error: 183334 HEMRC fault could not

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P104 : 10 Amp 700vdc Fuse - used in HEMRC and Axial Dynamic Brake and others See Long Description for more info 46-170021P104 10 Amp 700vdc F1 Fuse -used in HEMRC and Axial Dynamic Brake and others many pictures below Part #: 46-170021P104 Part Type: GP CONSUMABLES Part Weight: 11.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 10 AMPS 700 VOLTS FRU: Yes Status: Orderable Comments: FLS - FUSE 10 AMPS 700 VOLTS ULTRA FAST. USED ON FAST RHAPSODE 7/14/97 GWENN 08992 Fuse for Axial Brake Module Fuse for HEMRC Rotor controler (HSA-1.x) Fuse for HEMRC Rotor Control (2.x-4.x)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.