Symptom

Table will not drive in any direction. When table elevation buttons are pressed the gantry will tilt See Long Description When pressing table down the gantry tilts back. When pressing table up the gantry tilts forward. No errors logged on boot up or during a scan hardware reset. Gantry tilt buttons still work normally.

Error message

Error: 182606 Uncommanded Tilt motion Error: 182606 Host: ETC Process: Tilt File: unComMotion.c 1.16 3/26/92 Line: 55 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 50186/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 13:17:14:025 Log Series: 354 HW/SW error: Uncommanded Tilt motion. 182606 Uncommanded Tilt motion Error: 182606 Host: ETC Process: Tilt File: unComMotion.c 1.16 3/26/92 Line: 55 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 50186/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 13:17:14:025 Log Series: 354 HW/SW error: Uncommanded Tilt motion. Current Position 1054 (.01 deg) Last Stationary Position 1010 (.01 deg) Tolerance 25 (.01 deg)

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Pow

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296316P2 TABLE POWER SUPPLY See Long Description for Detail Part #: 46-296316P2 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: TABLE POWER SUPPLY FRU: Yes Status: Orderable Comments: FLS - 46-296316P1 REPLACED BY 46-296316P2 PER PCN 189228. 46-296316P2 IS BACKWARDS COMPATIBLE TO P1 PER WES HEIMAN. MS4636 4/5/95 24VDC @ 7.5A OUTPUT POWER SUPPLY. USED ON ZEUS TABLE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.