Pular para o conteúdo principal
GE Healthcare · CTHiSpeed24506565001105600124505997210

FastCals abort on the Fourth Cal

Symptom

FastCals abort on the Fourth Cal See Long Description with Errors: 97210, 245059 and 245065 Database error encountered due a corrupt sbc data base. Customer tried to scan a patient scout was ok but axial mode failed with Error: 97210 Database error encountered

Error message

Error: 245065 Aborting xrays during axial because scanComplete message to hdrSrv failed Error: 245065 Suite: CT01 Host: SBC0 Process: scanmgr File: do_axial_complete.m 1.16 Method: do_axial_complete() Line: 393 Function: Data Acquisition : SBC Processing Scan: 65001/11/4 Type: Xray-on Axial Scan Seq: 1056001 Aborting xrays during axial because scanComplete message to hdrSrv failed

Error: 245059 Error Encountered Communicating with Header Services Error: 245059 Suite: CT01 Host: SBC0 Process: scanmgr File: do_axial_complete.m 1.16 Method: do_axial_complete() Line: 387 Function: Data Acquisition : SBC Processing Scan: 65001/11/4 Type: Xray-on Axial Scan Seq: 1056001 Error Encountered Communicating with Header Services.

Error: 97210 Database error encountered See Long Description for complete error. Error: 97210 Suite: CT01 Host: SBC0 Process: hdrSrv File: ScnAlloc.m 1.24 Method: installScn: Line: 1759 Database error encountered for Exam 65001, Series 11, Scan 34. idbErr: (1044) Record key is defined in DB dbErr: (1510) Unable to install image

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues