Pular para o conteúdo principal
GE Healthcare · CTHiSpeed184163

HiSpeed CT/i 4.X with SGI Indigo Console, HiSpeed Advantage Gantry

Symptom

Various intermittent cradle drive related faults

Error message

Error: 184163 Timeout on helical gate signal (cradle sync). OR Timeout on Table gate signal (Cradle Sync Pulse). On a Lightspeed system the error may look like this: Fri Sep 8 11:48:58 2006 Error: 184163 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1526 Function: XRAY GENERATION : Table - Gantry Sync Scan: 5564/1/1 Type: Helical Exception Level: Pri/Most Severe Time: 11:48:58:255 Log Series: 132 Timeout on Table gate signal (Cradle Sync Pulse). HW/SW Error: Probable hardware, possible software, problem. (Timeout: 796). On a CTi system the error may look like this: Thu Feb 3 10:17:17 2005 Error: 184163 Suite: RP22 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.68 Method: No Method Line: 1351 Function: XRAY GENERATION : Table - Gantry Sync Scan: 4521/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 10:17:18:010 Log Series: 285 Timeout on helical gate signal (cradle sync). HW/SW Error: Probable hardware, possible software, problem.

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2236254 Axial Control Board - used in H2 gantry STC 2236254 Axial Control Board 2236254 Axial Control Board Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AXIAL CONTROL BOARD FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: COMMON TO MFG MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues