Pular para o conteúdo principal
GE Healthcare · MRExcite22538502549307

Collapsed image on Sagittal and Coronal scans. axial image looks normal but very thick.

Symptom

MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image

Error message

2253850 Gradient Link Hammer test failed. The GP received an invalid Z axis data value via the gradient data link. See long description for entire error. Sun Jun 19 00:24:18 2005 Error: 2253850 Host: lx-mr Process: diagCoord File: grad_hammer.c Line: 286 Diagnostic: Grad Hammer Iteration: 8 of infinite Action: 1 Task ID: 0x3d5c6c8 Time: 2549307 Object: SCP Test Task Gradient Link Hammer test failed. The GP received an invalid Z axis data value via the gradient data link. Data Exp = 000aaaaaH or 00055555H Data Rd = 000aaabaH Check the gradient data and clock fiber optic cable connections

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution