GE Healthcare · MRExcite22538502549307
Collapsed image on Sagittal and Coronal scans. axial image looks normal but very thick.
Symptom
MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image
Error message
2253850 Gradient Link Hammer test failed. The GP received an invalid Z axis data value via the gradient data link. See long description for entire error. Sun Jun 19 00:24:18 2005 Error: 2253850 Host: lx-mr Process: diagCoord File: grad_hammer.c Line: 286 Diagnostic: Grad Hammer Iteration: 8 of infinite Action: 1 Task ID: 0x3d5c6c8 Time: 2549307 Object: SCP Test Task Gradient Link Hammer test failed. The GP received an invalid Z axis data value via the gradient data link. Data Exp = 000aaaaaH or 00055555H Data Rd = 000aaabaH Check the gradient data and clock fiber optic cable connections
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.