GE Healthcare · CTLightSpeed260107705
LightSpeed 5.X (60kW)
Symptom
Gantry does not rotate. Customer noticed that gantry does not rotate.
Error message
Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. See LD for complete error Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Mon Mar 14 10:34:57 2005 Error: 260107705 Host: tgp Process: tgp_subsys File: scanbusDriver.c Line: 233 Mon Mar 14 10:34:57 2005 Host : Table Gantry Processor Ermes # : 260107705 Exception Class : Abort Severity : Secondary File : scanbusDriver.c Line# : 233 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Retries used: 0
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P101 - FUSE 20 AMPS 700 VOLTS - Fuse used in Axial Dynamic Brake and HEMRC (HSA - 1.X) and also in HEMRC (2.X-5.X) See LD for some of the locations of this fuse 46-170021P101 - FUSE 20 AMPS 700 VOLTS many pictures below 46-170021P101 - FUSE 20 AMPS 700 VOLTS - Fuse for Axial Dynamic Brake Part #: 46-170021P101 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 20 AMPS 700 VOLTS FRU: Yes Status: Orderable Comments: FLS - FUSE 20 AMPS 700 VOLTS USED ON FAST RHAPSODE GWENN 08992 7/2/97 Fuse for 5.X Axial Dynamic Brake part found in LIGHTSPEED 5.X ILLUSTRATED PARTS LIST DIRECTION 2349187-100, REVISION 14 Chapter 4 Gantry (Pro 16) Section 1 HP100 Vertical Assembly Table 4-17 Fuse for PRE 5.X Axial Dynamic Brake Fuse for HEMRC Rotor controler (HSA-1.X) Fuse for HEMRC Rotor Control (2.x-4.x)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required