GE Healthcare · MRExcite2244573

Intermittent DW-EPI failures with 2244573

Symptom

Intermittent DW-EPI failures with 2244573 Diffusion scans would intermittently fail to complete with 2244573 "The GP is reporting % Axis SGA Internal Wiring Fault and Internal Power Supply Undervoltage" In this case the failure was always Z axis only. Diffusion scans would intermittently fail to complete with 2244573 "The GP is reporting Z Axis SGA Internal Wiring Fault and Internal Power Supply Undervoltage" In this case the failure was always Z axis only.

Error message

2244573 The GP is reporting %c Axis SGA Internal Wiring Fault and Internal Power Supply Undervoltage 2244573 "The GP is reporting %c Axis SGA Internal Wiring Fault and\n\ Internal Power Supply Undervoltage. Perform a TPS Reset.\n\ Else, replace the SGA."

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.