Pular para o conteúdo principal
GE Healthcare · CTLightSpeed183182500014485818336550168

LightSpeed 3.X (Ultra, Plus, QXi) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 488 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 44858/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 18:11:57:355 Log Series: 807 I/O board pilot relays are closed. PDU pilot relay not energized. BCSTAT = 0 HVDC Rail voltage = 210V If the rail voltage is approximately 200 Volts, you are missing one phase and most likely, a blown fuse is the culprit. HVDC Rail voltage = somewhere near 100 V This in-between voltage probably indicates bad C1 C2 capacitors or 3 phase Bridge Rect. or other PDU Control Bd. probl

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2133533-3 CAPACITOR, 4600UF, 450V, 14A, 85DEGC Quantity 2 required See Long Description for details 2133533-3 Capacitor, 4600 uf, 450v, 85 deg, Quantity 2 required and replace both if you suspect either one has failed or causing problems. Part #: 2133533-3 Part Type: GP CONSUMABLES Part Weight: 2.27 LBR Repairable: No Returnable: No Hazardous Material: N Description: CAPACITOR, 4600UF, 450V, 14A, 85DEGC FRU: Yes Status: Orderable Parts Offered as: Current: 2133533-3 Harvested: Not Available Exchange: Not Available Comments: FLS - 4600DF, 450V, 14A, 85 DEGREE CU/O COMPACT PDU RHAPSODE CAN SOURCE FROM ON-LINE. MS4636 11/7/96
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.