GE Healthcare · CTHiSpeed186423
Will not scan with did not receive reply from JEDI
Symptom
Will not scan with did not receive reply from JEDI See Long Description Will not scan with Error: 186423 Code : 20-0014-02 :OGP did not receive any reply from JEDI in response to SysConfig CMD nor JEDI Capability Request in the error log
Error message
Error: 186423 OGP did not receive any reply from JEDI in response to SysConfig CMD nor JEDI Capability Request CMD. Code : 20-0014-02 Error: 186423 Host: CT02 Process: tgp_out_genetest File: tgp_errors.c Line: 365 XG Proc Comm Error. Code : 20-0014-02 Infomation detected by : OGP OGP did not receive any reply from JEDI in response to SysConfig CMD nor JEDI Capability Request CMD.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Component
- 5126997 AC/DC mid power assy (replaces 2214995) See long description for details Part "2214995" changed to "5126997 Part #: 5126997 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: AC-DC 3PH MID POWER ASSY_FRU FRU: Yes Status: Orderable Comments: Not Available Part #: 2214995 Part Type: GP CONSUMABLES Part Weight: 8.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: AC/DC 3PH MID POWER ASSY FRU: Yes Status: Orderable ====================== Parts Offered as: Current: 2214995 Harvested: Not Available Exchange: Not Available Comments: FLS - AC/DC 3PH MID POWER ASM U/O YMS PROSPEED LXI,DXI,FXI REPLACES 2193680 JEDI GENERATOR BRIDGE RECTIFIER=2121569 F1 FUSE : 2261172 ALSO USE ON CT/E GELS - DEFECTIVE PARTS TO BE PUT ON SHELVES FOR ANALYSIS PDB00/07/27 KANBAN CSS F00. 29-JUN-2000),POWER SUPPLY AC/DC 3PH MID
- Area
- CT
Full cause & step-by-step resolution