GE Healthcare · MRExcite2245100741345574386
ACGD Gradient Driver Test Fails.
Symptom
ACGD Gradient Driver Static Test Fails ACGD Gradient Driver Test Fails
Error message
2245100 GRADIENT DRIVER TESTS - Static Test Failed on % Axis Fri Jun 11 17:57:56 2004 Error: 2245100 Host: lx-mr Process: diagCoord File: grad_lib.c Line: 416 Diagnostic: GDT Stat&Dyn Iteration: 1 of 1 Action: 1 Task ID: 0xfc9990 Time: 741345 Object: SCP Test Task GRADIENT DRIVER TESTS - Static Test Failed on X A Fri Jun 2 17:37:01 2006 Error: 2245100 Host: lxmr3 Process: diagCoord File: grad_lib.c Line: 921 Diagnostic: GDT Static Iteration: 1 of 1 Action: 1 Task ID: 0x3d5c6c8 Time: 574386 Object: SCP Test Task GRADIENT DRIVER TESTS - Static Test Failed on the Z Axis. VBUS_700 had 2 failures out of 39. Ready Static Mode-> this happened on the z axis- tech swapped the power inputs and the error moved-ordred SGA power supply
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.