Pular para o conteúdo principal
GE Healthcare · MRSigna2225660

Unable to scan, Erbtec errors.

Symptom

Unable to scan with Erbtec errors

Error message

2225660 ERBTEC RF Amplifier Fault 81 3kV Arc Detected Thu Oct 23 19:32:24 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225660 File: PMH.m 1.29 Method: spi_rf_mon Line: 3035 ERBTEC RF Amplifier fault 81. 3kV arc detected. Possible defective IPA tube or bad processor board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues