HART Log May Incorrectly Report SCP Board When Actually SCP2 Board Installed

Symptom

HART Log May Incorrectly Report SCP Board When Actually SCP2 Board Installed FE noticed that SCP Board was reported in HARdware Tracker (HART) Log of 14.0 system. FE knew that 14.0 only supports SCP2 Board. FE obtained SCP2 Board and swapped out old board but HART Log still reported SCP Board. The hexidecimal identifier code that HART uses to differentiate an SCP2 from an SCP Board was recently changed by the board vendor. As a result an SCP2 Board may be incorrectly reported in the HART Log as an SCP Board. This issue does not affect scanner performance. (See Resolution concept for the complete SMILe released 06/13/2006) FE noticed that SCP Board was reported in HARdware Tracker (HART) Log of 14.0 system. FE knew that 14.0 only supports SCP2 Board. FE obtained SCP2 Board and swapped out old board but HART Log still reported SCP Board. The hexidecimal identifier code that HART uses to differentiate an SCP2 from an SCP Board was recently changed by the board vendor. As a result an SCP2 Board may be incorrectly reported in the HART Log as an SCP Board. This issue does not affect scanner performance.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR